Errors & Messages
|
Solutions
|
|
When starting the system, power indicator light is not lit.
|
1)Check the power supply cable where on the rear of equipment whether connected already or not;
2)Check fuse, if find the fuse are burned as power supply sudden change, use spare fuse of the same mode (250V 10A) to replace the fuse.
|
|
When starting the system, keyboard light is lit, monitor has output but no ultrasonic image.
|
1)Check the power supply cord of the monitor, whether connected already or not;
2)Check the on-off knob of monitor, whether opened up or not;
3)Check the VGA cable whether connected to the VGA 1 of equipment rear or not.
|
|
When starting the system, the monitor output with format, appears “No probe, please connect probe” dialog box.
|
Switch off power, and check the connection of probe.
|
|
The video printer cannot print when press the [print] key on the big panel
|
1)Check whether the video printer electrified or not;
2)Whether there is printer paper or not in the printer;
3) Check whether the video printer well connected to the system or not.
|
|
Image has interference and the image at high depth is not clear
|
1)Move and avoid interference;
2)Use separate 220V power cord;
3)Ground the main unit well;
4)Adjust depth parameter of image.
|
|
Appear “Ultrasound Scanner” dialog box on the screen.
|
The power supply is not stable. Use a voltage stabilizer.
|
|
The date and time on the screen is not correct.
|
Press [SETUP] key , select [System] and set the correct time and date.
|